Journal of Clinical Microbiology, January 1998, p. 203-206, Vol. 36, No. 1
0095-1137/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
Evaluation of the OptiMAL Test for Rapid Diagnosis
of Plasmodium vivax and Plasmodium
falciparum Malaria
Carol J.
Palmer,*
John F.
Lindo,
Winslow I.
Klaskala,
Jose A.
Quesada,
Rina
Kaminsky,
Marianna K.
Baum, and
Arba L.
Ager
Center for Disease Prevention, University of
Miami School of Medicine, Miami, Florida, and Escula University
Hospital, Tegucigalpa, Honduras
Received 11 July 1997/Returned for modification 19 August
1997/Accepted 6 October 1997
 |
ABSTRACT |
The development of rapid and specific diagnostic tests to identify
individuals infected with malaria is of paramount importance in efforts
to control the severe public health impact of this disease. This study
evaluated the ability of a newly developed rapid malaria diagnostic
test, OptiMAL (Flow Inc., Portland, Oreg.), to detect Plasmodium
vivax and Plasmodium falciparum malaria during an
outbreak in Honduras. OptiMAL is a rapid (10-min) malaria detection test which utilizes a dipstick coated with monoclonal antibodies against the intracellular metabolic enzyme parasite lactate
dehydrogenase (pLDH). Differentiation of malaria parasites is based on
antigenic differences between the pLDH isoforms. Since pLDH is produced only by live Plasmodium parasites, this test has the
ability to differentiate live from dead organisms. Results from the
OptiMAL test were compared to those obtained by reading 100 fields of traditional Giemsa-stained thick-smear blood films. Whole-blood samples
were obtained from 202 patients suspected of having malaria. A total of
96 samples (48%) were positive by blood films, while 91 (45%) were
positive by the OptiMAL test. The blood films indicated that 82% (79 of 96) of the patients were positive for P. vivax and 18%
(17 of 96) were infected with P. falciparum. The OptiMAL test showed that 81% (74 of 91) were positive for P. vivax
and 19% (17 of 91) were positive for P. falciparum. These
results demonstrated that the OptiMAL test had sensitivities of 94 and 88% and specificities of 100 and 99%, respectively, when compared to
traditional blood films for the detection of P. vivax and
P. falciparum malaria. Blood samples not identified by
OptiMAL as malaria positive normally contained parasites at
concentrations of less than 100/µl of blood. Samples found to contain
P. falciparum were further tested by two other commercially
available rapid malaria diagnostic tests, ParaSight-F (Becton
Dickinson, Cockeysville, Md.) and ICT Malaria P.f. (ICT Diagnostics,
Sydney, Australia), both of which detect only P. falciparum. Only 11 of the 17 (65%) P. falciparum-positive blood samples were identified by the ICT and
ParaSight-F tests. Thus, OptiMAL correctly identified P. falciparum malaria parasites in patient blood samples more often
than did the other two commercially available diagnostic tests and
showed an excellent correlation with traditional blood films in the
identification of both P. vivax malaria and P. falciparum malaria. We conclude that the OptiMAL test is an
effective tool for the rapid diagnosis of malaria.
 |
INTRODUCTION |
Malaria has had a resurgence in many
tropical areas (6). The disease now occurs in more than 90 countries worldwide, and it is estimated that there are over 500 million clinical cases and 2.7 million malaria-caused deaths per year
(8). A multitude of factors have contributed to the
reemergence of malaria, including (i) insecticide resistance in the
Anopheles mosquito, (ii) social instability resulting in
movements of unexposed nonimmune individuals into areas where malaria
is endemic, and (iii) the failure to develop an effective malaria
vaccine (6, 15). Compounding the problems of malaria's
geographical expansion and of increasing morbidity and mortality are
the emergence and rapid spread of antimalarial-drug resistance (2,
7), which necessitate the use of more expensive and sometimes
toxic antimalarial drugs and longer treatment courses (9).
In addition, the cyclic recurrence of malaria epidemics has a
tremendous impact on the health infrastructure in developing countries
and adversely affects local economies, since infected individuals are
often too debilitated to work (6).
One of the most pronounced problems in controlling the morbidity and
mortality caused by malaria is limited access to effective diagnosis
and treatment in areas where malaria is endemic (10). Clinical diagnosis of infection with the malaria parasite requires microscopic observation of parasites on a Giemsa-stained blood smear.
This technique has undergone little improvement since its development
at the beginning of this century. Diagnosis of malaria by this method
can be problematic, since it requires up to 60 min of preparation time,
is labor-intensive, and requires considerable expertise for its
interpretation, particularly at low levels of parasitemia
(3). In addition, in patients with Plasmodium
falciparum malaria, the parasites can be sequestered and are not
always present in peripheral blood. Thus, a P. falciparum
infection could be easily missed due to the absence of the parasite in
a blood film.
Recently, a new rapid malaria detection test, OptiMAL (Flow Inc.,
Portland, Oreg.), was introduced. This test is based on detection of an
enzyme produced by live parasites, parasite lactate dehydrogenase
(pLDH), and has the ability to differentiate the four major
Plasmodium species associated with human malaria
(Plasmodium vivax, P. falciparum,
Plasmodium ovale, and Plasmodium malariae) in
under 10 min. A trial evaluation of the OptiMAL test was designed to
assess its effectiveness in differentiating malaria parasites in
patient blood samples during a malaria outbreak in Honduras.
 |
MATERIALS AND METHODS |
Study site.
This study was conducted in the Trujillo area of
northern Honduras during a malaria outbreak in January 1997. Permission
to conduct blood testing was obtained from the Honduran Ministry of
Health. The majority of the blood samples were collected from individuals living in rural areas and small villages. A total of 202 whole-blood samples were collected from patients with malaria-like symptoms, including fever and/or chills, of several days' duration.
Sample collection.
Informed consent was obtained from all
patients. Five milliliters of venous blood was drawn into EDTA-coated
syringes, distributed into sterile test tubes, and placed immediately
on ice. Thin- and thick-smear blood films were made on-site at the time
of specimen collection. Whole-blood samples plus the thin- and
thick-smear microscope slides were transported the next day to the
University of Miami School of Medicine laboratories for analysis.
According to the local treatment guidelines for malaria, symptomatic
patients were treated immediately with chloroquine (500 mg daily for 4 days) by local health care staff prior to diagnosis. Due to the lengthy
ongoing malaria outbreak, this was the only antimalarial medication
available at that time. Patients were instructed to return in 4 days
for additional chloroquine if clinical signs persisted.
Malaria diagnosis with thin- and thick-smear films.
Thin-
and thick-smear blood films were stained with 2% Giemsa stain in
phosphate-buffered saline (pH 7.2) and analyzed under the microscope at
×1,000 magnification for the presence of malaria parasites. The
average time spent per slide varied depending on the number of
parasites present in the sample. The study was blinded, since results
from microscope slides were not shared with the individuals performing
the OptiMAL test until all samples were processed. Slides were read
independently by two experts. Parasitemia levels were calculated with
results from thick-smear films. Parasites were counted in 100 consecutive fields. Parasite densities were calculated by assuming 0.2 µl of blood per thick smear and that each microliter of blood
contained 10,000 leukocytes (11, 14).
Malaria diagnosis with OptiMAL.
All whole-blood samples were
tested with the OptiMAL assay. This test utilizes a dipstick coated
with monoclonal antibodies against the intracellular metabolic enzyme
pLDH. The pLDH antigen is present in and released from
parasite-infected erythrocytes. Differentiation of malaria parasites is
based on antigenic differences between the pLDH isoforms. Since pLDH is
produced only by live Plasmodium parasites, this test has
the ability to differentiate live from dead organisms. The test uses
two reagents (A and B) included in the kit. Normally, the test is
performed in the field by obtaining blood from a patient by a finger
stick. The test was not yet available when we left for Honduras, but we
obtained and evaluated it upon our return. Briefly, 1 drop of blood was mixed with 2 drops of reagent A, and the sample was allowed to migrate
to the top of the OptiMAL dipstick. After 8 min, the OptiMAL strip was
cleared by adding 2 drops of reagent B. The appearance of a dark band
on the strip indicates a positive reaction for any one of the four
major malarial species infecting humans. The monoclonal antibody
attached at this area of the strip is against an enzyme common to the
four target Plasmodium species. If P. falciparum
was present in the test sample, a second band appeared on the strip.
The monoclonal antibody at this site is specific for P. falciparum only. A mixed infection with P. falciparum
and another Plasmodium species is indicated when both genus-
and species-specific bands appear and the genus-specific band is much
darker and more intense than the species-specific band. The darker band
is present with mixed infections, since all stages of the parasite are
present in the blood during P. vivax infections, while only
two stages (ring and gametocyte) are present in the peripheral blood
during P. falciparum infections. Since the pLDH enzyme is
present in all stages and since there are more stages present during
P. vivax infections, there is a stronger response with the
OptiMAL stick. A positive control band appears on each strip as an
indicator that the test is working correctly (Fig.
1).
P. falciparum malaria diagnosis with the ParaSight-F
and ICT malaria tests.
Two commercially available tests,
ParaSight-F (Becton Dickinson, Cockeysville, Md.) and ICT Malaria P.f.
(ICT Diagnostics, Sydney, Australia), were tested on blood samples that
were positive for P. falciparum to determine the level of
agreement with the OptiMAL test and blood smears. The manufacturer's
instructions were followed.
 |
RESULTS |
Comparison of OptiMAL and blood film methods.
A total of 202 blood samples were tested for malaria parasites by the OptiMAL method,
and the results were compared to results obtained from reading thin-
and thick-smear blood films. The blood film results indicated that 48%
(96 of 202) of the patients were infected with malaria, based on the
morphologies of the parasite stages. Among the positive patient
samples, P. vivax was present in 82% (79 of 96) while
P. falciparum was present in 18% (17 of 96).
Correspondingly, the OptiMAL test results indicated that 45% (91 of
202) of the patient samples were positive for malaria parasites.
Infections with P. vivax accounted for 81% (74 of 91) of
the positive samples, while infections with P. falciparum
accounted for 19% (17 of 91) of the total malaria cases. Both methods
identified one patient with a mixed infection of P. falciparum and P. vivax. The mixed infection has been
tabulated with the P. falciparum infection numbers for ease
of analysis.
The blood films identified five
P. vivax-positive samples
that were not identified by the OptiMAL test; however, there was
100%
agreement between blood film results and OptiMAL results
for the other
74 samples containing
P. vivax. Although both methods
detected 17 cases of
P. falciparum infection, there were 2 cases
detected by OptiMAL that were not detected by the blood films
and
2 cases detected by blood film that were not detected by the
OptiMAL
method. OptiMAL had sensitivities of 94% (95% confidence
interval
[CI], 85.2 to 97.6%) and 88% (95% CI, 62.3 to 97.9%)
and
specificities of 100% (95% CI, 96.2 to 100.01%) and 99% (95%
CI,
95.5 to 99.8%), respectively, when compared to traditional
blood films
for the detection of
P. vivax and
P. falciparum
infections
(Table
1). Positive and
negative predictive values were 100%
(95% CI, 93.9 to 100.0%) and
96% (95% CI, 90.7 to 98.6%), respectively,
for
P. vivax
and 88% (95% CI, 62.3 to 97.9%) and 99% (95% CI,
95.7 to 99.8%),
respectively, for
P. falciparum.
Parasitemia.
Parasitemia levels ranged from 0.001 to 12% in
the patient samples. This translated into 50 to 600,000 parasites/ml of
blood. Patients with P. vivax present in the blood film and
who were diagnosed as negative by the OptiMAL method had parasitemia
values ranging from 0.001 to 0.008%. The two patients with P. falciparum present in blood films that was not detected by OptiMAL
had parasitemia values of 0.001 and 0.004%. However, parasitemia
levels of 0.001% in other blood samples were detected by OptiMAL,
indicating the level of error that is inherent in the enumeration of
organisms on blood films.
The OptiMAL test proved more sensitive at higher levels of parasitemia
(Table
2). Sensitivity in detecting
P. vivax was 40%
when parasites were present at less than
100/µl of blood and rose
to 100% when parasites were present at more
than 100/µl. Two samples
containing
P. vivax at more than
100/µl were missed by OptiMAL.
Sensitivity of detection for
P. falciparum was 67% when parasites
were present at less than
100/µl of blood, and one
P. falciparum-positive
blood
sample was missed by OptiMAL at parasite levels greater
than 100/µl
of blood.
Comparison of OptiMAL, ParaSight-F, and ICT Malaria P.f. for the
detection of P. falciparum.
Results from the ParaSight-F
test and the ICT test did not always agree with results from the
OptiMAL test in the diagnosis of P. falciparum malaria. Only
11 of the 17 (65%) P. falciparum-positive blood samples
were identified by the other two tests. For these 11 positive samples,
there were differences between the tests, since the ParaSight-F test
identified one positive sample missed by the ICT test and the ICT test
identified one positive sample missed by the ParaSight-F test. Five of
the six samples that were not judged positive by the ParaSight-F and
ICT tests were from patients with 50 P. falciparum
organisms/µl of blood (0.001% parasitemia). The sixth was from a
patient with 0.003% parasitemia.
 |
DISCUSSION |
This study compared the diagnosis of malaria by a new rapid test,
OptiMAL, with malaria diagnosis by traditional microscopy and found
that the two methods yielded comparable results. A total of 202 symptomatic patients were tested; blood films identified 48% of these
as positive for malaria parasites, while the OptiMAL test identified
45% as malaria positive. That some malaria infections detected by
blood films were not detected by the OptiMAL test may be explained by
the fact that OptiMAL detects pLDH, which is produced only by living
parasites. It is possible that some of the patients infected with
malaria medicated themselves when malaria symptoms appeared during this
outbreak and did not report this to the attending clinician. There are
several possible explanations for discrepancies in test results
obtained by blood film examination and by the OptiMAL test, including
(i) insufficient detection of low parasitemia levels by OptiMAL, (ii)
the fact that OptiMAL detects only live parasites producing pLDH, (iii)
the sequestration of parasites, and (iv) false-positive reactions.
Thus, the P. vivax and P. falciparum organisms
observed in the seven blood film samples judged negative by OptiMAL may
have been dead and not yet cleared from the host. The two patient blood
samples in which OptiMAL detected P. falciparum and whose
blood films were negative may be explained by the fact that P. falciparum can sequester and not be present in circulating blood.
This suggests that the OptiMAL test may provide a more precise
diagnosis of patients infected with P. falciparum malaria by
detecting parasites that would be missed by traditional blood film
screening. Another explanation could be that the patient blood samples
contained parasites at concentrations below the OptiMAL test's
detection levels.
Currently there are two other commercially available rapid tests, the
ParaSight-F and ICT tests, for diagnosis of P. falciparum malaria. While both of these tests offer diagnosis of malaria in under
15 min, there are differences between these two tests and the OptiMAL
test evaluated in this study. The first difference is that the ICT and
ParaSight-F tests are based on detection of parasite histidine-rich
protein 2, which has been shown to remain in the blood at least 28 days
after the initiation of antimalarial therapy (5, 13). Thus,
parasite clearance after drug therapy may not be determined and
patients who have cleared the parasite may be falsely diagnosed as
malaria positive and given additional drug therapy. The second
difference is that the ICT and ParaSight-F tests detect only one of the
four Plasmodium species known to infect humans, P. falciparum. Since over half of the patient samples in this study
contained P. vivax, the use of the ParaSight-F and ICT
Malaria P.f. tests alone would have meant that the majority of the
patients in this study would not have been diagnosed with malaria.
The sensitivities of the three tests were not similar in our study,
since OptiMAL identified more P. falciparum-positive samples than did the ICT and ParaSight-F tests. There are few published studies
on the ICT test. A report by Garcia et al. (4) indicated that the ICT test was able to detect P. falciparum malaria
with parasite densities exceeding 80/µl but that it was unknown
whether the test worked with densities of less than 80/µl. There are
several reports in the literature on the ParaSight-F test. Its
sensitivity has been reported by other researchers to range from 40 to
100% with parasite densities under 1,000/µl (1, 5, 12);
however, those authors used different methods for the determination of parasite density. This is a common problem, since different
laboratories base parasite densities on leukocyte counts which vary
between 1,000 and 10,000/µl of blood. Other laboratories determine
parasite density by using erythrocyte formulations. The use of various methods to determine levels of malaria parasitemia is turning into a
major problem, as it makes comparisons of test results from different
malaria researchers difficult.
There are compelling reasons to justify the implementation of a rapid
malaria diagnostic test in the field. None of the rural clinics in this
study had the ability to diagnose malaria on-site due to a lack of
microscopes and trained technicians to evaluate blood films. Moreover,
patient follow-up is difficult due to economic constraints. Many people
cannot afford transportation, so they walk several hours, some carrying
small children, to reach a local clinic; once they have been seen, they
do not return. Diagnosis must therefore be immediate in order to
provide proper treatment. This is of particular importance in areas
such as Honduras, where dengue fever and other tropical diseases mimic
each other's symptoms (e.g., fever, chills, and headache). In this
study, roughly half of the patients symptomatic for malaria were
actually infected with malaria parasites. However, all received
antimalarial medication, based on symptoms alone.
The world health care community is in desperate need of rapid and
precise diagnostic tools to assist in controlling malaria, monitoring
drug therapy, and limiting mortality and suffering. Our data
demonstrated that the OptiMAL test is an effective, sensitive, and
rapid diagnostic test for malaria that could be invaluable in the fight
to control malaria. Additional studies to evaluate the use of OptiMAL
in monitoring drug therapy and in the detection of drug-resistant
malarial strains are under way.
 |
ACKNOWLEDGMENTS |
This project was funded by NIH Fogarty grant 5D43TW00017-08.
We thank the Honduran Ministry of Health for allowing access to malaria
patients and Honduran local health care staff for their assistance with
this study. We are most grateful to Merita Aviles and Xiaofeng Liang
for technical assistance.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: University of
Miami School of Medicine, Center for Disease Prevention, 1400 NW 10th Ave., 10th Floor (D21), Miami, FL 33136. Phone: (305) 243-3221. Fax:
(305) 243-4687.
 |
REFERENCES |
| 1.
|
Beadle, C.,
G. W. Long,
W. R. Weiss,
P. D. McElroy,
S. Maret,
A. J. Oloo, and S. L. Hoffman.
1994.
Diagnosis of malaria by detection of Plasmodium falciparum HRP-2 antigen with a rapid dipstick antigen-capture assay.
Lancet
343:564-568[Medline].
|
| 2.
|
Collins, W. E., and G. M. Jefferey.
1996.
Primaquine resistance in Plasmodium vivax.
Am. J. Trop. Med. Hyg.
55:243-249.
|
| 3.
|
Dourado, H.,
N. Abdon, and S. J. Martins.
1994.
Falciparum malaria.
Infect. Dis. Clin. N. Am.
8:207-223[Medline].
|
| 4.
|
Garcia, M.,
S. Kirimoama,
D. Marlborough,
J. Leafasia, and K. H. Rieckmann.
1996.
Immunochromatographic test for malaria diagnosis.
Lancet
347:1549[Medline].
|
| 5.
|
Humar, A.,
C. Ohrt,
M. A. Harrington,
D. Pillai, and K. C. Kain.
1997.
ParaSight F test compared with the polymerase chain reaction and microscopy for the diagnosis of Plasmodium falciparum malaria in travelers.
Am. J. Trop. Med. Hyg.
56:44-48.
|
| 6.
|
Krogstad, D. J.
1996.
Malaria as a reemerging disease.
Epidemiol. Rev.
18:77-89[Free Full Text].
|
| 7.
|
Longworth, D. L.
1995.
Drug-resistant malaria in children and travelers.
Antimicrob. Resistance Pediatr.
42:649-664.
|
| 8.
|
Nussenzweig, R. S., and F. Zavala.
1997.
A malaria vaccine based on a sporozoite antigen.
N. Engl. J. Med.
336:128-130[Free Full Text].
|
| 9.
|
Olliaro, P.,
J. Cattani, and D. Wirth.
1996.
Malaria, the submerged disease.
JAMA
275:230-233[Abstract/Free Full Text].
|
| 10.
|
Pan American Health Organization.
1996.
Malaria in the Americas.
PAHO Bull.
17:1-8.
|
| 11.
|
Petersen, E.,
N. T. Marbiah,
A. New, and A. Gottschau.
1996.
Comparison of two methods for enumerating malaria parasites in thick blood films.
Am. J. Trop. Med. Hyg.
55:485-489.
|
| 12.
|
Shiff, C. J.,
J. Minjas, and Z. Premji.
1994.
The Parasight-F test: a simple rapid manual dipstick test to detect Plasmodium falciparum infection.
Parasitol. Today
10:494-495.
|
| 13.
|
Verle, P.,
L. N. Binh,
T. T. Lieu,
P. T. Yen, and M. Coosemans.
1996.
ParaSight-F test to diagnose malaria in hypo-endemic and epidemic prone regions of Vietnam.
Trop. Med. Int. Health
1:794-796.
|
| 14.
|
Warhurst, D. C., and J. E. Williams.
1996.
Laboratory diagnosis of malaria.
J. Clin. Pathol.
49:533-538[Free Full Text].
|
| 15.
|
Wernsdorfer, W. H.
1994.
Epidemiology of drug resistance in malaria.
Acta Trop.
56:143-156[Medline].
|
Journal of Clinical Microbiology, January 1998, p. 203-206, Vol. 36, No. 1
0095-1137/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
Fernando, S. D., Abeyasinghe, R. R., Galappaththy, G. N. L., Rajapaksa, L. C.
(2009). Absence of Asymptomatic Malaria Infections in Previously High Endemic Areas of Sri Lanka. Am J Trop Med Hyg
81: 763-767
[Abstract]
[Full Text]
-
de Oliveira, A. M., Skarbinski, J., Ouma, P. O., Kariuki, S., Barnwell, J. W., Otieno, K., Onyona, P., Causer, L. M., Laserson, K. F., Akhwale, W. S., Slutsker, L., Hamel, M.
(2009). Performance of Malaria Rapid Diagnostic Tests as Part of Routine Malaria Case Management in Kenya. Am J Trop Med Hyg
80: 470-474
[Abstract]
[Full Text]
-
Lee, S. W., Jeon, K., Jeon, B. R., Park, I.
(2008). Rapid Diagnosis of Vivax Malaria by the SD Bioline Malaria Antigen Test When Thrombocytopenia Is Present. J. Clin. Microbiol.
46: 939-942
[Abstract]
[Full Text]
-
RATSIMBASOA, A., RANDRIAMANANTENA, A., RAHERINJAFY, R., RASOARILALAO, N., MENARD, D.
(2007). WHICH MALARIA RAPID TEST FOR MADAGASCAR? FIELD AND LABORATORY EVALUATION OF THREE TESTS AND EXPERT MICROSCOPY OF SAMPLES FROM SUSPECTED MALARIA PATIENTS IN MADAGASCAR. Am J Trop Med Hyg
76: 481-485
[Abstract]
[Full Text]
-
VAN DEN BROEK, I., HILL, O., GORDILLO, F., ANGARITA, B., HAMADE, P., COUNIHAN, H., GUTHMANN, J.-P.
(2006). EVALUATION OF THREE RAPID TESTS FOR DIAGNOSIS OF P. FALCIPARUM AND P. VIVAX MALARIA IN COLOMBIA. Am J Trop Med Hyg
75: 1209-1215
[Abstract]
[Full Text]
-
Lee, N., Baker, J., Andrews, K. T., Gatton, M. L., Bell, D., Cheng, Q., McCarthy, J.
(2006). Effect of Sequence Variation in Plasmodium falciparum Histidine- Rich Protein 2 on Binding of Specific Monoclonal Antibodies: Implications for Rapid Diagnostic Tests for Malaria.. J. Clin. Microbiol.
44: 2773-2778
[Abstract]
[Full Text]
-
Chilton, D, Malik, A N J, Armstrong, M, Kettelhut, M, Parker-Williams, J, Chiodini, P L
(2006). Use of rapid diagnostic tests for diagnosis of malaria in the UK. J. Clin. Pathol.
59: 862-866
[Abstract]
[Full Text]
-
ZIMMERMAN, P. A., THOMSON, J. M., FUJIOKA, H., COLLINS, W. E., ZBOROWSKI, M.
(2006). DIAGNOSIS OF MALARIA BY MAGNETIC DEPOSITION MICROSCOPY. Am J Trop Med Hyg
74: 568-572
[Abstract]
[Full Text]
-
SINTASATH, D. M., GHEBREMESKEL, T., LYNCH, M., KLEINAU, E., BRETAS, G., SHILILU, J., BRANTLY, E., GRAVES, P. M., BEIER, J. C.
(2005). MALARIA PREVALENCE AND ASSOCIATED RISK FACTORS IN ERITREA. Am J Trop Med Hyg
72: 682-687
[Abstract]
[Full Text]
-
Ndao, M., Bandyayera, E., Kokoskin, E., Gyorkos, T. W., MacLean, J. D., Ward, B. J.
(2004). Comparison of Blood Smear, Antigen Detection, and Nested-PCR Methods for Screening Refugees from Regions Where Malaria Is Endemic after a Malaria Outbreak in Quebec, Canada. J. Clin. Microbiol.
42: 2694-2700
[Abstract]
[Full Text]
-
Perandin, F., Manca, N., Calderaro, A., Piccolo, G., Galati, L., Ricci, L., Medici, M. C., Arcangeletti, M. C., Snounou, G., Dettori, G., Chezzi, C.
(2004). Development of a Real-Time PCR Assay for Detection of Plasmodium falciparum, Plasmodium vivax, and Plasmodium ovale for Routine Clinical Diagnosis. J. Clin. Microbiol.
42: 1214-1219
[Abstract]
[Full Text]
-
Palmer, C. J., Bonilla, J. A., Bruckner, D. A., Barnett, E. D., Miller, N. S., Haseeb, M. A., Masci, J. R., Stauffer, W. M.
(2003). Multicenter Study To Evaluate the OptiMAL Test for Rapid Diagnosis of Malaria in U.S. Hospitals. J. Clin. Microbiol.
41: 5178-5182
[Abstract]
[Full Text]
-
Forney, J. R., Wongsrichanalai, C., Magill, A. J., Craig, L. G., Sirichaisinthop, J., Bautista, C. T., Miller, R. S., Ockenhouse, C. F., Kester, K. E., Aronson, N. E., Andersen, E. M., Quino-Ascurra, H. A., Vidal, C., Moran, K. A., Murray, C. K., DeWitt, C. C., Heppner, D. G., Kain, K. C., Ballou, W. R., Gasser, R. A. Jr.
(2003). Devices for Rapid Diagnosis of Malaria: Evaluation of Prototype Assays That Detect Plasmodium falciparum Histidine-Rich Protein 2 and a Plasmodium vivax-Specific Antigen. J. Clin. Microbiol.
41: 2358-2366
[Abstract]
[Full Text]
-
IQBAL, J., MUNEER, A., KHALID, N., AHMED, M. A.
(2003). PERFORMANCE OF THE OPTIMAL TEST FOR MALARIA DIAGNOSIS AMONG SUSPECTED MALARIA PATIENTS AT THE RURAL HEALTH CENTERS. Am J Trop Med Hyg
68: 624-628
[Abstract]
[Full Text]
-
Richardson, D. C., Ciach, M., Zhong, K. J. Y., Crandall, I., Kain, K. C.
(2002). Evaluation of the Makromed Dipstick Assay versus PCR for Diagnosis of Plasmodium falciparum Malaria in Returned Travelers. J. Clin. Microbiol.
40: 4528-4530
[Abstract]
[Full Text]
-
Iqbal, J., Khalid, N., Hira, P. R.
(2002). Comparison of Two Commercial Assays with Expert Microscopy for Confirmation of Symptomatically Diagnosed Malaria. J. Clin. Microbiol.
40: 4675-4678
[Abstract]
[Full Text]
-
Playford, E. G., Walker, J.
(2002). Evaluation of the ICT Malaria P.f/P.v and the OptiMal Rapid Diagnostic Tests for Malaria in Febrile Returned Travellers. J. Clin. Microbiol.
40: 4166-4171
[Abstract]
[Full Text]
-
Moody, A.
(2002). Rapid Diagnostic Tests for Malaria Parasites. Clin. Microbiol. Rev.
15: 66-78
[Abstract]
[Full Text]
-
Forney, J. R., Magill, A. J., Wongsrichanalai, C., Sirichaisinthop, J., Bautista, C. T., Heppner, D. G., Miller, R. S., Ockenhouse, C. F., Gubanov, A., Shafer, R., DeWitt, C. C., Quino-Ascurra, H. A., Kester, K. E., Kain, K. C., Walsh, D. S., Ballou, W. R., Gasser, R. A. Jr.
(2001). Malaria Rapid Diagnostic Devices: Performance Characteristics of the ParaSight F Device Determined in a Multisite Field Study. J. Clin. Microbiol.
39: 2884-2890
[Abstract]
[Full Text]
-
Iqbal, J., Sher, A., Rab, A.
(2000). Plasmodium falciparum Histidine-Rich Protein 2-Based Immunocapture Diagnostic Assay for Malaria: Cross-Reactivity with Rheumatoid Factors. J. Clin. Microbiol.
38: 1184-1186
[Abstract]
[Full Text]
-
Iqbal, J., Sher, A., Hira, P. R., Al-Owaish, R.
(1999). Comparison of the OptiMAL Test with PCR for Diagnosis of Malaria in Immigrants. J. Clin. Microbiol.
37: 3644-3646
[Abstract]
[Full Text]
-
Grobusch, M. P., Jelinek, T., Hänscheid;, T., Mishra, B., Samantaray, J. C., Kumar, A., Mirdha, B. R.
(1999). False Positivity of Rapid Antigen Detection Tests for Diagnosis of Plasmodium falciparum Malaria: Issue Appears To Be More Complicated than Presented. J. Clin. Microbiol.
37: 3781-3782
[Full Text]
-
Tjitra, E., Suprianto, S., Dyer, M., Currie, B. J., Anstey, N. M.
(1999). Field Evaluation of the ICT Malaria P.f/P.v Immunochromatographic Test for Detection of Plasmodium falciparum and Plasmodium vivax in Patients with a Presumptive Clinical Diagnosis of Malaria in Eastern Indonesia. J. Clin. Microbiol.
37: 2412-2417
[Abstract]
[Full Text]
-
Jelinek, T., Grobusch, M. P., Schwenke, S., Steidl, S., von Sonnenburg, F., Nothdurft, H. D., Klein, E., Löscher, T.
(1999). Sensitivity and Specificity of Dipstick Tests for Rapid Diagnosis of Malaria in Nonimmune Travelers. J. Clin. Microbiol.
37: 721-723
[Abstract]
[Full Text]