Journal of Clinical Microbiology, March 2008, p. 869-875, Vol. 46, No. 3
0095-1137/08/$08.00+0 doi:10.1128/JCM.01424-07
Copyright © 2008, American Society for Microbiology. All Rights Reserved.

Divisions of Microbiology and Cytopathology, Department of Pathology and Laboratory Medicine, and the George Burns and Gracie Allen Research Institute, Cedars-Sinai Medical Center, Los Angeles, California 90048
Received 16 July 2007/ Returned for modification 30 October 2007/ Accepted 21 December 2007
| ABSTRACT |
|---|
|
|
|---|
| INTRODUCTION |
|---|
|
|
|---|
HPV typing and colposcopy are ancillary tests for patients with an abnormal cervical cytology, while cervical biopsy for the demonstration of CIN3 remains the only diagnostic test for cervical neoplasia. Colposcopy is invasive and expensive and can miss one-third of CIN3 lesions (8). Although cervical cancer is induced by HPV, the different HPV types are not equally carcinogenic. Fifteen oncogenic or high-risk (HR) strains are implicated in cervical carcinogenesis. Of these HR HPV types, HPV type 16 (HPV 16) and HPV 18 are etiologically associated with CIN3 and invasive cervical carcinomas (CIN lesions of grade 3 or higher) in 60% and 10 to 20% of specimens, respectively (17). The ASCUS LSIL Triage Study Group reported that the 2-year cumulative absolute risk for CIN lesions of grade 3 or higher is 32.5% for HPV 16-positive ASCUS specimens (2), which raises a clinical urgency. Moreover, the persistence of HPV infection is most often associated with HPV 16 and 18 (9, 14, 24, 26), suggesting that the identification of these two HPV types may play an important role in the risk stratification of patients with ASCUS for appropriate monitoring and, possibly, heightened follow-up.
Approximately 50% of ASCUS specimens demonstrate HR HPV infections (7). Algorithmically, the need for clinical follow-up for ASCUS is considerably intensified by a positive screen for HR HPV. Most contemporary HPV tests that indicate positive results for HPV do not yet specifically identify HPV 16 and 18, nor do they indicate the level of inherent patient risk when the assay results are positive. As one approach to the recently recognized need for a level of specificity for HPV 16 and 18 in cervical cytology specimens (2, 12), we examined a new and yet undescribed HPV screening assay, the second-generation Invader HPV (Inv2) assay (Third Wave Technologies, Inc., Madison, WI). The Inv2 assay screens for HPV DNA and subgroups the HR HPV types with three separate HR HPV-specific probe sets.
The Inv2 assay is available as analyte-specific reagents (ASRs) and provides a potentially semiautomated, high-throughput system for the detection of 14 HR HPV types. According to the manufacturer of the Inv2 assay ASRs, the Inv2 assay is similar to the first-generation Invader HPV HR assay but with the addition of a probe for HR HPV 66 (Table 1). The Inv2 assay HR HPV-specific probe sets include A5/A6 (HPV 51, 56, and 66), A7 (HPV 18, 39, 45, 59, and 68), and A9 (HPV 16, 31, 33, 35, 52, and 58). Positivity for probes A9 and A7 is suggestive of the presence of HPV 16 and 18, respectively (Table 1). The Inv2 assay features an internal positive control through the detection of the human histone-2, or H2be, gene, ensuring the quality of the DNA and informative results for specimens with negative assay results.
|
| MATERIALS AND METHODS |
|---|
|
|
|---|
Specimens in SurePath medium were vortexed at 3,000 rpm for 20 s, subsequently enriched by dispensing approximately 8 ml of the specimen over 4 ml of PrepMate density reagent, and then centrifuged at 800 x g for 10 min. The supernatant was decanted, and the pellet was vortexed. Probe B of the HC2 test (Digene) was used for the postenrichment sample, with residual specimen stored at 4°C for a period up to 3 months prior to the Inv2 and amplification assays.
NucliSens easyMAG total nucleic acid extraction. Total nucleic acids were extracted from the specimens in SurePath medium by using the NucliSens easyMAG platform, a second-generation system for the automated isolation of nucleic acid from clinical samples that is based on the silica extraction technology (bioMérieux, Durham, NC). One to 2 ml of the sample in SurePath medium was centrifuged at 1,150 x g for 15 min. The supernatant was removed, and the cell pellet was resuspended in 400 µl of Genfind lysis buffer (Agencourt Bioscience Corporation, Beverly, MA). This mixture was incubated at 99°C for 10 min and then cooled to ambient (room) temperature. Nine microliters of proteinase K (Agencourt Bioscience Corporation) was added to each tube containing the reaction mixture, and the tubes were incubated at 37°C for 30 min on a thermomixer. Following incubation, the DNA was extracted by use of the NucliSens easyMAG platform and eluted to a final volume of 110 µl.
HPV typing by Inv2 assay. The extracted DNAs were tested for the presence of HR HPV DNAs in a laboratory-developed HPV assay with the Inv2 ASRs, including HPV Oligos (version 2.0; Third Wave Technologies, Inc.). Three Inv2 assay HPV reaction mixtures representing Inv2 assay oligonucleotides (probe sets) A5/A6, A7, and A9, respectively, were prepared and contained the specific oligonucleotides, the enzyme cleavase, MgCl2, and oligonucleotides specific for the human H2be gene.
Ten microliters of each sample or control was added to three separate wells, which each contained one of the targets for the three HPV-specific probe sets, in a 96-well reaction plate. Twenty microliters of Chill-Out liquid wax (Bio-Rad, Hercules, CA) was added to each well. After incubation at 95°C for 5 min and cooling to 63°C, 10 µl of each Inv2 assay reaction mixture was added to the appropriate specimens and controls. The reaction plate was incubated at 63°C for 4 h and then read in a GENios reader (Tecan Systems, Inc., San Jose, CA) with the following settings: for the 6-carboxyfluorescein (FAM) signal, excitation at 485 nm and emission at 535 nm; for the red signal, excitation at 560 nm and emission at 612 nm. Data analysis and HR HPV detection were performed with a Microsoft Excel software-based HPV (version 2.0) Invader data analysis worksheet (Third Wave Technologies, Inc.).
Analytic criteria for Inv2 assay results and cutoff values. The fluorescent signals of all specimens were expressed as the FAM (HPV specific) and red (H2be-specific) signals. The levels of the red and FAM signals over zero (red fold over zero [Red FOZ] and FAM fold over zero [FAM FOZ]) were calculated for each of the three probe sets as follows: FAM FOZ = FAM signal of sample/FAM signal of no-DNA target solution and Red FOZ = red signal of sample/red signal of no-DNA target solution.
The FAM FOZ ratio was subsequently calculated with the FAM FOZ values from each of the three probe sets for a given sample (see below). When the lowest FAM FOZ value was <1, the value was set equal to 1.0 for the calculation of the FAM FOZ ratio: FAM FOZ ratio = highest FAM FOZ value/lowest FAM FOZ value and FAM FOZ ratio = middle FAM FOZ value/lowest FAM FOZ value.
Specimens with FAM FOZ ratios
1.4 were considered positive for HR HPV. When the FAM FOZ ratio was <1.4 but the specimen demonstrated a FAM FOZ ratio of
1.5 with all three HPV-specific probe sets, the specimen was called multiple positive, or the presence of "three fluorescence signals" (3FS).
An average Red FOZ value was calculated for each specimen and served as an indicator for genomic DNA levels. A negative reaction for HPV for specimens with average Red FOZ values <1.5 was considered insufficient for a result.
PCR-LA. The linear array HPV genotyping assay (PCR-LA; Roche Molecular Systems, Inc., Branchburg, NJ) employs the amplification of target DNA by PCR and nucleic acid hybridization and has been designed to detect 37 cervical HPV DNA genotypes, including 13 HR types (HPV 16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, and 68). PCR-LA is for research use only (Fig. 1).
|
| RESULTS |
|---|
|
|
|---|
For determination of the reproducibility of the Inv2 assay, the coefficients of variation were established through repeat testing for high to low viral "loads" in dilution assays, through replicate testing in separate runs, on different days. CaSki and HeLa cell DNAs, which were the source materials for HPV 16 and HPV 18, respectively, were employed for known viral "copy" numbers (Advanced Biotechnology, Inc., Columbia, MD). The reaction mixtures contained HPV 16 (1.2 x 103 to 10 x 103 copies in four dilutions) or HPV 18 (5 x 103 to 10 x 103 copies in two dilutions). For HPV 16 and 18, the coefficients of variation for the FAM FOZ ratios ranged from 6.63 to 2.53% and from 9.33 to 4.72%, respectively (range from low to high viral loads). The FAM FOZ ratios were reproduced with good precision.
All specimens were screened for the presence of HPV by the Inv2 and HC2 assays, and the results were considered within the context of ASCUS or negative cytology results (Tables 2 to 5).
|
|
Twelve ASCUS specimens had discordant results and comprised seven Inv2 assay-negative and HC2 assay-positive specimens and five Inv2 assay-positive and HC2 assay-negative specimens, all of which underwent PCR-LA testing (Tables 4 and 5). Of the seven Inv2 assay-negative and HC2 assay-positive samples, two specimens were found to contain HR HPV types, including a sample with HPV 35 and 56 and a second sample with HPV 16 and 39. Among the five specimens that were Inv2 assay positive and HC2 assay negative, three samples had HR HPV types, including a specimen with both HPV 16 and HPV 51, a specimen with HPV 39, and a specimen with HPV 66. Of all the discordant specimens, PCR-LA detected HPV 16 in 2 of the 12 samples: 1 specimen that was Inv2 assay positive and HC2 assay negative and 1 specimen that was Inv2 assay negative and HC2 assay positive, with the Inv2 and HC2 assays each failing to detect a sample in which HPV 16 was detected by PCR-LA (Fig. 1). For samples with an ASCUS cytology, the concordance between the Inv2 and HC2 tests was 86.8% (79/91 samples).
|
|
Overall, the results of the Inv2 assay demonstrated a concordance of 86.6% compared to the results of the HC2 assay. By use of PCR-LA as the reference assay for discordant specimens, the Inv2 assay had an overall sensitivity of 95.6% and a specificity of 96.1%, a positive predictive value of 93.6%, and a negative predictive value of 97.4%.
All of the specimens with an ASCUS result and all specimens with a negative cytology result were reevaluated in the context of the Inv2 assay results. No cytomorphological differences were seen among the specimens, whether they were positive, negative, or indeterminate for HPV (Fig. 2 and 3).
|
|
| DISCUSSION |
|---|
|
|
|---|
This study describes the performance characteristics of a new HPV test, the second-generation Inv2 assay. Clinically, the Inv2 test detects 14 oncogenic HPV types and segregates the viruses into three groups according to the results obtained with separate probe sets, thus allowing the selection of women for whom further subtyping is needed. In this study, cervical samples were submitted in SurePath ethanol-based transport medium, with DNA extractions for Inv2 testing obtained by semiautomated methods. SurePath medium has been increasingly popular as a liquid fixative for gynecologic cytology; moreover, screening for HPV from a liquid fixative is rapidly growing as an important testing modality for cancer prevention. For the Inv2 assay, H2be serves as an internal control and ensures that informative DNA is extracted from the specimen. In this study, the extractions for the Inv2 assay were "clean"; that is, the semiautomated extraction resulted in a low level of extraneous background DNA and provided quality specimen DNA.
In examining the performance characteristics of the Inv2 assay with specimens with an ASCUS cytology, the results obtained by the Inv2 assay were compared to those obtained by the HC2 assay. The discordances between the two tests were examined by a third method, PCR-LA. For ASCUS specimens, the results obtained by the Inv2 assay demonstrated a good correlation (86.8%) with those obtained by screening by the HC2 assay. The Inv2 test demonstrated that 48.4% (44/91) of ASCUS specimens were positive for HR HPV, with 38.4% (35/91) representing potential HPV types 16 and 18. This incidence of HR HPV correlates with that of 45.5 to 59% published previously for females with an ASCUS cytology with HR HPV (7).
Otherwise, 12 ASCUS specimens were discordant for HPV typing by the Inv2 assay and the HC2 assay. With PCR-LA used as a reference, for specimens with an ASCUS cytology, the rates of false-positive results for HR HPV were 4.6% (2/44) for the Inv2 assay and 10.9% (5/46) for the HC2 assay. The rates of false-negative results were 4.3% (2/47) for the Inv2 assay and 6.7% (3/45) HC2 assay. The results of the Inv2 assay were falsely negative for HR HPV 16, 35, 39, and 56 in two specimens, and the results of the HC2 assay were falsely negative for HPV 16, 39, 51, and 66 in three specimens. Two ASCUS cases screened negative for HPV, one each by the Inv2 assay and the HC2 assay. This finding is notable, since each specimen was positive for HPV 16 by PCR-LA.
HPV typing of 39 specimens with negative cytology results demonstrated HR HPV by the Inv2 assay (with probe set A9 or A7) in 13.9% (5/36) of specimens, and 4 (11.1%) of theses specimens potentially involved HPV 16 and/or 18. These findings correlate with the previously published findings that 17% of patients with a negative cytology were infected with HR HPV (7). Specimens with a negative cytology screened as discordant for HPV by the Inv2 and HC2 assays were also examined by PCR-LA. Of the three specimens tested, PCR-LA detected an HR HPV type (type 56) in a single specimen that was Inv2 assay positive.
Overall, the results of the Inv2 assay demonstrated a concordance of 86.6% compared to the results of the HC2 assay. When PCR-LA was used as a reference for discordant specimens, the Inv2 assay had an overall sensitivity and specificity of 96%, a positive predictive value of 94%, and a negative predictive value of 97%.
Six specimens (three specimens with an ASCUS cytology and three specimens with a negative cytology result) demonstrated 3FS, which was classified as an indeterminate result in this study. All of these specimens were negative by the HC2 assay. All specimens had FAM FOZ values
4.0. PCR-LA was performed with the three specimens with an ASCUS cytology and an indeterminate result and demonstrated HPV 59 in one specimen. The significance of indeterminate results is not clear at present. Additional specimens with indeterminate results will need to be studied. It is possible that the cutoff values for the Inv2 test will need to be further optimized in the future. Alternatively, for the SurePath fixative, the specimen DNA concentration and/or the duration of fixation may require standardization to prevent indeterminate results. Importantly, no differences in cytological morphology were seen among the specimens that were positive, negative or indeterminate for HPV.
The specificity of assays for HPV subtypes is now a major concern, namely, because patients may be subjected to unnecessary stress and procedures secondary to false-positive results for HR types. The HC2 HPV screening assay detects at least 15 HPV genotypes not included in its HR probe set (19). It may be argued that it is important to detect HPV with sensitivity, irrespective of specificity; however, this may not attend to the caveats of risk and the economic cost associated with a false-positive test result. Specificity for HR HPV and the associated clinical implications—the need for close follow-up, including colposcopy—are relevant issues which suggest that a high sensitivity and a high specificity are paramount to clinical acumen. The Inv2 and HC2 tests appear to demonstrate similar sensitivities, although the Inv2 test may offer an improvement in specificity (19). From the results of the present study, the Inv2 test offers a reliable means of grouping of HR HPV types, which is potentially of advantage in triaging of patients for follow-up for possible HPV 16 and 18 infections. We have semiautomated the extraction phase for Inv2 testing for high throughput. Otherwise, the LA-PCR allows the specific recognition of individual types within both low-risk and HR HPV groups, although the test remains unavailable for clinical use at this time. In its present format, the application of PCR-LA would be problematic (because of strip hybridization) for high throughput.
In characterizing the Inv2 assay, the assay appears to be simple to perform, is amenable to use with the SurePath fixative, may be semiautomated for DNA extraction, and shows a good sensitivity and a good specificity for ASCUS cytology cases. For routine surveillance for HPV, the Inv2 test appears to be potentially useful as a first-line test for screening for HPV. The Inv2 test demonstrates few false-positive results without compromising the sensitivity of detection of critical HPV types, including HPV 16 and 18. Patients positive by the Inv2 test with the A9 and/or A7 probe set may warrant further close follow-up and/or testing for possible HPV 16 and 18 infection.
In the future, new modalities for testing for HPV should address HPV types 16 and 18. However, on the basis of even a limited association with invasive cervical disease (28), because of the obvious clinical stakes, it will remain prudent to continue to screen for multiple HR HPV types, such as for ASCUS cytology cases, as well as for additional triage for HPV types 16 and 18.